Sterile Heat Exchangers 

heater-coler-consumables-spectrum-medical 

 

Quantum Perfusion range of sterile disposable heat exchangers

The Quantum Perfusion range of sterile disposable heat exchangers have been designed by Spectrum Medical to provide the safest solution for minimizing the growth environment for the Mycobacterium Chimaera bacteria.

Working exclusively with the Quantum Heater-Cooler technology and its Glycol heat transfer fluid. Both the Arterial heat exchanger and the Cardiopleagia heat exchanger are compact in design, with low priming volumes and minimal pressure drops.

Technical Specifications

Standard Heat Exchanger: High Flow (adult size)

  • Priming volume blood side: 55 ml
  • Exchange Surface: 0.12 m2
  • Priming volume HTF side: 70 ml
  • Maximum flow rate: 8 l/min

Standard Heat Exchanger: Medium Flow (pediatric size)

  • Priming volume blood side: 22 ml
  • Exchange Surface: 0.05 m2
  • Priming volume HTF side: 38 ml
  • Maximum flow rate: 3 l/min
 

Standard Heat Exchanger: Low Flow (neonatal size)

  • Priming volume blood side: 11 ml
  • Exchange Surface: 0.028 m2
  • Priming volume HTF side: 22 ml
  • Maximum flow rate: 1 l/min

Cardioplegia Heat Exchanger

  • Priming volume blood side: 22 ml
  • Exchange Surface: 0.028 m2
  • Priming volume HTF side: 22 ml
  • Maximum flow rate: 0.5 l/min
  • Relief Valve Opening Pressure: 500 mmHg